Cardiac output measurement
Press Cardiac Output Mode
· position patient supine and flat
· check monitor programmed with correct: height, weight, Co-efficient and syringe size
· check that the thermistor cable is connected
· attach the injectate system to the proximal lumen and ensure that an adequate volume of 5% dextrose is available for Cardiac output measurements
v enter the following values into monitor
v Mean P.A.P. and A.P.
v P.C.W.P.
v C.V.P.
· Perform measurement
· draw up 10ml of 5% dextrose
· activate CO start button on peak inspiration and inject solution within 2 seconds
· inspect CO curve, looking for smooth rapid upstroke to peak and gradual down slope to baseline
· repeat CO procedure obtaining 3 measurements that are within 10% of the median valve
· Press HAEMO DATA STORE key to store C. O. values
·
|
·
|
![]() |
Tracing shows cardiac rhythm (top), intraaortic counterpulsation waveform (middle) and PAP and PCWP trace (bottom)