Wednesday, September 16, 2009

Ventilators

Ventilators

Star Infrasonic

Draeger Evita 4

Star Infrasonic ventilator

Classification:

· microprocessor controlled electronic ventilator

· multiple weaning modes

· three power sources (Gas, electrical, internal and external battery source)

Modes available:

· volume control (SIMV + PPS, ASSIST CONTROL, CPAP & PPS )

· pressure control Time cycled (SIMV + PPS, ASSIST CONTROL, CPAP & PPS)

· pressure control I:E cycled (SIMV + PPS, ASSIST CONTROL, CPAP & PPS)

· BIPAP

Quick test: Mandatory prior to initiating ventilation

· depress and hold down visual reset and alarm silence buttons while turning the panel ON/OFF switch to ON

· after 5 secs the performance, safety and diagnostic screen will be displayed

· move the cursor to quick test and press enter

· follow prompts (there are fourteen tests that are automatically performed during the quick test; if a test fails, use the cursor to rerun this test, once you’ve rectified the problem)

Default ventilator setting Volume control SIMV 10 BPM x TV (10 mls per kg) x

FIO 2 .5 x Peep 5cms x PPS 20 cms Flow LPM 60 x decelerating waveform.

Select mode Volume control

SIMV + PS

Settings:

· set tidal volume at 800 mls

· set the intermittent mandatory ventilator rate at 8

· set flow rate at 60 LPM

· set trigger sensitivity at 2 cm

· set pause 0.5 sec

· set pressure support at 20 cm H 2 O

· peep 5 cm H 2 O

· FiO 2 .5 until adequacy of ventilation checked by blood gas analysis

· decelerating waveform

· sigh rate 0

· apnoea rate 8

Alarm limits:

· high inspiratory pressure limit 35 cm H 2 O

· low inspiratory pressure limit 10 cm H 2 O

· low mechanical volume 400 mls

· low spontaneous volume 0

· low minute volume 4L

· low PEEP/CPAP 2cm

· high respiratory rate 26 bpm

· apnoea 20 secs

Measurement of Auto Peep

· autopeep must be measured if inverse I:E ratio (in controlled modes only)

· select screen change and enter

· toggle to respiratory mechanics and enter

· toggle to Autopeep and enter Autopeep will be measured over three breaths

Measurement of NIF (must be > 20 cm H 2 O)

· NIF must be measured prior to extubation (in spontaneous modes only)

· select screen change and enter

· toggle to respiratory mechanics and enter

· toggle to NIF and enter

· enter while instructing patient to breath in

· only repeat if <> cm H2 O as very uncomfortable for the patient

Measurement of SVC (must be > 10 mls/kg)

· SVC must be measured prior to extubation (in spontaneous modes only)

· select screen change and enter

· toggle to respiratory mechanics and enter

· toggle to SVC and enter

· enter while instructing patient to breath out

PRESSURE CONTROL I:E CYCLED

· select pressure control I:E cycled and enter

· select assist control and enter

· select I:E 1:2 and enter

· select rate 8-12 bpm and enter

· select flow 160 LPM and enter

· select Oxygen 50 % and enter

· select PEEP/CPAP 5cm and enter

· select Sensitivity at 2 cm and enter

· select apnoea rate 8 bpm and enter.

· toggle down to right hand corner of the screen and select start

The Pressure control mode will now be enabled.

PRESSURE CONTROL TIME-CYCLED

· select pressure control Time-cycled and enter

· toggle to select (assist control or SIMV + PPS or CPAP + PPS) and enter

· select PCV pressure at 10-20 cms H 2 O or equivalent to plateau pressure in SIMV mode

· select insp. time at 1.5 and enter

· toggle to select rate 8-12 bpm and enter

· toggle to select flow 160 LPM and enter

· toggle to select oxygen 50 % and enter

· toggle to select PEEP/CPAP 5cm and enter

· toggle to select Sensitivity at 2 cm and enter

· toggle to select PPS 0-20 cms H 2 O

· select apnoea rate 8 bpm and enter

· toggle down to right hand corner of the screen and select start

The Pressure control mode will now be enabled.

CPAP +PPS ( BIPAP )

Weaning/assist mode.

· select volume control

· select CPAP + PPS

· set tidal volume at 10 mls/kg

· set the intermittent mandatory ventilator rate at 8-12

· set flow rate at 60 LPM

· set pause 0.5 sec

· set pressure support at 20 cm H 2 O

· Peep 5 cm H 2 O

· FiO 2 .5 until adequacy of ventilation checked by blood gas analysis

· decelerating waveform

· sigh rate 0

· apnoea rate 6